OPERATION

TRIGGER MENU

-9

dBm

-50

TTL

GSM BURST

FROM GENERATOR

BURST TRIGGER SYNC

FROM GENERATOR

Figure 4-3.Typical Arming Diagram

 

TYPE

 

[TRGTYP

 

GTTYP]

 

LEVEL

NOTE

[TRGLVL

GTLVL]

 

Effective range is to ap-

 

proximately –30 dBm

 

and is only active in DC

Trig 1

ranges 1 and 2.

 

 

Trig 2

 

 

Trig 1&2

1.Connect to the rear panel digital input.

2.Select TriggerSetupARMINGBlanking ON.

3.Set the polarity of the blanking (System menu) Example power meter reading: –9.16 dBm.

The Type selection (RISE or FALL) sets the trigger for a rising or falling edge. When the trigger source is set to INTA or INTB (Inter- nal A or B) the power meter triggers on a power level which is rising or falling.

The Level selection sets the internal trigger level. When the trigger source is set to either INTA or INTB (internal sensor A or B) the channel triggers on a power level (in dBm) given by the sensor. This value must not take any cal factors or offsets that the meter applies into account.

If Trigger Channel 1 SOURCE is set to Manual, this softkey initiates a measurement for channel 1.

If Trigger Channel 2 SOURCE is set to Manual, this softkey initiates a measurement for channel 2.

If Trigger Channels 1 and 2 SOURCE are set to Manual, this softkey triggers both channels simultaneously.

ML2430A OM