GPIB OPERATION | ML24XXA NATIVE COMMANDS |
binary data:
Same data as that recieved by CFURD
Remarks: Loads binary data into the cal factor table. This command will automatically save the data to the sensor.
CFUPT | Preset cal factor table | SENSOR |
Syntax: CFUPT <s>,<table number>
s:A or B
table
number: 1 to number of tables supported by the sensor type
Remarks: Presets the cal factor table to the factory settings. The preset table is auto- matically saved to the sensor. For a universal sensor, separate cal factor tables exist for option 1 (when fitted) and normal operation. This function will preset the table relating to the currently selected mode.
CFURD | Cal factor table binary read | SENSOR |
Syntax: CFURD <s>,<table number>
s:A or B
table
number: 1 to number of tables supported by the sensor type F for the factory default table
Remarks: This command outputs the cal factor table in binary mode in the following form:
CFURD<space><length of binary data>,<binary data>
<length of binary data>: Total length of the binary data message, in bytes, after the comma.
<binary data>: Made up of :
a. 8 bytes; 7 for the identity, plus a NULL terminator
b. 2 bytes representing the number of table pair entries
c. The cal factor table data in binary form. The binary data is output in en- tries which are frequency/factor pairs of six bytes. The frequency is held in
To convert these into real numbers the first four bytes of an entry are read into a LONG variable, cast to a float and then divided by
ML2430A OM | |
|
|