ML24XXA NATIVE COMMANDS | GPIB OPERATION |
and measurement channel setups.
When in Readout mode, the Channel 1 and Channel 2 values, and the max/min values if present, are printed below the header.
In Profile and Power vs. Time modes, a graph is printed out below the header with all the details shown.
PRNSEL | Select the type of printer | SYSTEM |
Syntax: PRNSEL <type>
type: HP340 BJC80
Remarks: Available printer selections are the HP DeskJet 340 and Canon BJC80. Other 300, 500, 600 Series and later HP printers are typically compatible. If the Canon BJC80 printer is selected, it must be set to EPSON LQ emu- lation mode for proper operation. Refer to the printer manual for instruc- tions on setting the emulation mode.
Query: PRNSEL?
Returned
String: PRNSEL <printer>
RCD | Range Calibrator data request | DATA OUTPUT |
Syntax: RCD <s>
s:A or B
Remarks: Returns the results from an ML2419A Range Calibrator run after the Range Calibrator is disconnected from the power meter. While still connected to the Range Calibrator, the results can be printed but not read via GPIB. The results include values for each end of each sensor range and the zero value, and are kept in
Response format:
'RCD<ws><sensor>,<state>[,<zero value>,<range 1 upper>,<range 1 lower>,........]'
<state>:
If state is FALSE, no data follows because there are no valid results for this sensor available.
If state is TRUE, the results for the selected sensor are displayed in the fol- lowing order:
zero value, range 1 upper, range 1 lower, range 2 upper, range 2 lower,
ML2430A OM | |
|
|