Anritsu ML2430A operation manual System, Specifications

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 223
Image 223

 

 

Appendix A

 

 

Specifications

A-1

INTRODUCTION

This appendix provides system specifications for the ML2430A Series

 

 

Power Meters along with listings of system defaults and error messages.

A-2

SYSTEM

This section provides overall system specifications.

 

SPECIFICATIONS

 

 

 

 

Frequency

 

 

 

Range:

10 MHz to 90 GHz (sensor dependent)

 

 

Power

Meter specifications apply to MA2400A Series Power

 

 

Sensors:

Sensors. Compatible with MA and MP Series sensors.

 

 

Sensor

MA2420A/B Series Fast Thermal Sensors: 50 dB

 

 

Dynamic

MA2440A Series High Accuracy Power Sensors: 87 dB

 

 

Range:

CW

 

 

 

MA2470A Series Power Sensors: 90 dB CW

 

 

 

MA2460A/B Series Fast diode sensors: 80dB

 

 

 

MA2480A Series Universal Power Sensor: 80 dB

 

 

Power

 

 

 

Measurement

–70 to +47 dBm (0.1 nW to 50W), sensor/attenuator

 

 

Range:

dependent. Use couplers for higher power levels.

 

 

Voltage Meas-

 

 

 

urement

 

 

 

Range:

0.00 to 20.00 V, nominal

 

 

Display Range:

 

 

 

 

–99.999 to +99.999 dB

 

 

Display Reso-

Selectable from 0.1 dB to 0.001 dB limited to 0.01 dB

 

 

lution:

in graphical display modes; Linear power units, 3 to 6

 

 

 

digit, 1 – 3 digits selectable to right of decimal nW – W;

 

 

 

Voltage, 1 – 2 digits selectable to right of decimal.

 

 

Offset Range:

–99.99 to +99.99 dB. Fixed value or frequency de-

 

 

 

pendent table.

Display Units:

dBm, dB, dBr, dBmV, dB￿V, W, %, Volts

ML2430A OM

A-1

 

 

Page 223