GPIB OPERATION | ML24XXA NATIVE COMMANDS |
Remarks: Places a single ASCII character '1' on the GPIB output queue when the conditions for the *OPC command are met. An operation is complete when all input messages before the command have been completed and any responses have been read out of the output buffer.
Example: RGH A, 1; RGH B, 2; *OPC?
Returns a ‘1’ on the GPIB output when it has finished setting the range hold commands.
*RCL | Recall a stored setup | SYSTEM |
Syntax: *RCL <val>
val: 1 to 10
Remarks: The ML2430A Series can store up to 10 instrument configurations for con- venient recall. The configuration parameters stored are:
Sensor Setup, Channel Setup, and Trigger Setup.
This command sets the ML2430A Series to a configuration previously stored in memory locations 1 through 10. Trying to recall a setup that has not been saved will set the execution bit in the event register (EXE in ESR).
Related
Commands: *SAV
*RST | Reset Device | GPIB 488.2 |
Syntax: *RST
Remarks: Resets the ML2430A Series to the default configuration (see Appendix A, Section
*SAV | Save configuration | SYSTEM |
Syntax: *SAV <val>
val: 1 to 10
Remarks: Saves the configuration of the power meter into the memory location specified. Sensor Setup, Channel Setup, and Trigger Setup are saved along with all other instrument parameters.
ML2430A OM | |
|
|