ML24XXA NATIVE COMMANDS

GPIB OPERATION

GPRST

Reset min/max graph

PROFILE SETUP

Syntax: GPRST

Remarks: When profile DATA HOLD mode is set to Min/Max points, this command is used to reset the min/max values for each data point.

Related

Commands: GRPIX

GRAUTO Auto scaling

SYSTEM

Syntax: GRAUTO

Remarks: Auto scale for all graphic modes (Profile, Source Sweep and Power vs. Time). This command auto scales the y axis only based on the currently displayed data.

GRAVG

Average between profile cursors

PROFILE SETUP

Syntax: GRAVG <state>

state: ON or OFF

Remarks: Turns on or off averaging between cursors. The data returned by the GRDRQ command includes the average of all data points between the cur- sors if GRAVG is turned ON.

Related

Commands: GRDRQ

Query: GRAVG?

Returned

String: GRAVG <state>

GRCP

Connect points on profile

PROFILE SETUP

Syntax: GRCP <state>

state: ON or OFF

Remarks: When set to ON, creates a line graph by connecting the profile data points together. The default is ON.

Query: GRCP?

6-40

ML2430A OM

 

 

Page 127
Image 127
Anritsu ML2430A Syntax Gprst, Related Commands Grpix, Syntax Grauto, Related Commands Grdrq Query GRAVG?, Query GRCP?