ML24XXA NATIVE COMMANDS | GPIB OPERATION |
power
value: power
Remarks: Determines the start power level of power sweep mode.
Query: SRCSTPWR?
Returned
String: SRCSTPWR <power>
START | Initial startup self test command | GPIB SETUP |
Syntax: START
Remarks: This is useful for ATE control. After the system has been given time to start up, this command can be used to find out what state the system is in. If the self test has failed, 'CONT' can be used to get the system running. This is an initial startup self test status command and will return one of the follow- ing:
0 - Passed self test and running.
1 - Startup self test running.
In this stage of the startup process, all commands except STERR, START, CONT and GPIB 488.2 event and status commands will produce a GPIB execution error. STERR will return the selftest result string.
Related
Commands: STERR, CONT
STATUS | Status Message | DATA OUTPUT |
Syntax: STATUS
Remarks: Replies with the power meter’s current state code. In this format, the num- ber of letters specifies the number of digits, with preceding zeroes for pad- ding. The format is:
’ABCDEFGHIJKLMNNOOPQRRRRSSSSTUVWXYZ12’
where: A = Operating mode: '0' = Digital readout, '1' = Profile mode channel 1, '2' = Profile mode channel 2, '3' = Power vs. Time channel 1, '4'
=Power vs. Time channel 2, ‘5’ = Source sweep chan. 1, ‘6’ = Source sweep chan. 2..
B = Channel 1 input configuration: '0' = OFF, '1' = A, '2' = B, '3' =
C = Channel 2 input configuration: '0' = OFF, '1' = A, '2' = B, '3' =