HP 438A EMULATION COMMANDS | GPIB OPERATION |
Related
Commands: DD
DR | Set GPIB address |
Syntax: DR<val>
val: 1 to 30
Remarks: Changes the device address when operating in HP emulation mode. The power meter default address is 13.
EMUL | GPIB emulation mode |
Syntax: EMUL <mode>
mode: ML24XX (Anritsu ML2430A Series native mode) HP436A
HP437B
HP438A
Remarks: Set the GPIB emulation to emulate other types of power meters. This com- mand is available in any emulation mode, and resets the whole GPIB inter- face when the emulation mode is changed.
When selecting GPIB emulation modes, the instrument configures itself to the preset conditions of the instrument to be emulated. For example, when selecting HP 438A emulation, the front panel menus pass through the pre- sets for the HP 437B (which presets sensor A to dBm) then selects HP 438A emulation (which presets sensor A to Watts).
FA | auto average |
Syntax: FA
Remarks: Automatic Filter on. Allows the system to automatically select the filter used to reduce the jitter in the display.
Related
Commands: FM, FH
FH | average hold |
Syntax: FH
ML2430A OM | |
|
|