Anritsu ML2430A operation manual Related Commands DD, Syntax DRval

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 207
Image 207

HP 438A EMULATION COMMANDS

GPIB OPERATION

Related

Commands: DD

DR

Set GPIB address

Syntax: DR<val>

val: 1 to 30

Remarks: Changes the device address when operating in HP emulation mode. The power meter default address is 13.

EMUL

GPIB emulation mode

Syntax: EMUL <mode>

mode: ML24XX (Anritsu ML2430A Series native mode) HP436A (Hewlett-Packard)

HP437B (Hewlett-Packard)

HP438A (Hewlett-Packard) ML4803 (Anritsu ML4803A Series)

Remarks: Set the GPIB emulation to emulate other types of power meters. This com- mand is available in any emulation mode, and resets the whole GPIB inter- face when the emulation mode is changed.

When selecting GPIB emulation modes, the instrument configures itself to the preset conditions of the instrument to be emulated. For example, when selecting HP 438A emulation, the front panel menus pass through the pre- sets for the HP 437B (which presets sensor A to dBm) then selects HP 438A emulation (which presets sensor A to Watts).

FA

auto average

Syntax: FA

Remarks: Automatic Filter on. Allows the system to automatically select the filter used to reduce the jitter in the display.

Related

Commands: FM, FH

FH

average hold

Syntax: FH

6-120

ML2430A OM

 

 

Page 207
Image 207
Anritsu ML2430A operation manual Related Commands DD, Syntax DRval