GPIB OPERATION

ML24XXA NATIVE COMMANDS

NOTE

If the buffering enabled is set to OFF and '*OPC?' is used, the '*OPC?' will clear the output buffer of any previous re- sponse data so only the '1' will appear.

CAL

Cal sensor to 0 dBm reference

CALIBRATION

Syntax: CAL <s>

s:A or B

Remarks: Performs a 0dBm calibration when the sensor is attached to the reference 0 dBm source on the ML2430A Series (or another 0 dBm reference source). If the calibration fails, the 'execution error' bit in the Event Status Register is set.

CFADJ

Cal Adjust

SENSOR

Syntax: CFADJ <s>, <units>, <val>

s:A or B

units: %, PCT, DB, or DBM val: .07 to 150%

+31.55 to -1.76dB

Remarks: Sets a calibration factor to be used when performing a 0 dBm calibration and the calibration factor source is set to 'Manual.' This value is the only factor applied when performing a 0 dBm calibration. If the sensor calibra- tion factor source is set to V/GHz or Frequency, the sensor internal EE- PROM correction value at 50 MHz is used.

Examples: CFADJ A, %, 99

Sets the calibration factor to 99% for sensor A.

CFADJ A, DB, 0.2

Sets the calibration factor to 0.2dB for sensor A.

Query: CFADJ? <s>

Returned

String: CFADJ <s>,<units>,<val>

CFCAL

Cal factor manual setting

SENSOR

Syntax: CFCAL <s>, <units>, <val>

s:A or B

units: %, PCT, DB, or DBM

ML2430A OM

6-21

 

 

Page 108
Image 108
Anritsu ML2430A operation manual Syntax CAL s, Cfadj, Cfcal