ML24XXA NATIVE COMMANDS

GPIB OPERATION

This message can be manipulated to program a different table using the

CFULD command.

CFUSAV Cal factor table save

SENSOR

Syntax: CFUSAV

Remarks: This command saves the cal factor table currently being edited to the ap- propriate sensor. Processing may take a couple of seconds. Any command that can select a new sensor and/or cal factor table for changing, will not automatically save any previous changes made. It is the users responsibil- ity to issue a CFUSAV command.

CFUSEL

Select cal factor table

SENSOR

Syntax: CFUSEL <s>,<table number>

s:A or B

table

number: table number or combination to use 0 = factory default table

1 to 10 = user table being used

11 to 20 = factory table + user table being used

Remarks: Selects the cal factor table or combination of tables to be used and auto- matically updates the sensor.

Example: CFUSEL A,13

Selects the factory table plus user table 3 in sensor A.

CFUTBL

Number of cal factor tables in the sensor

SENSOR

Syntax: CFUTBL <s>

s:A or B

Remarks: Returns the number of cal factor tables available in the selected sensor.

CFUUSE Number of cal factor table being used

SENSOR

Syntax: CFUUSE <s>

s:A or B

6-26

ML2430A OM

 

 

Page 113
Image 113