GPIB OPERATION

GPIB PC CARD SETUP

6-5GPIB PC CARD

The following GPIB driver configuration set up is recommended for reliable GPIB

SETUP

communication with the ML2430A Series power meter. The set up is expressed

 

in the terms used by the National Instruments GPIB ISA and PCI cards and driv-

 

ers for WIN95 and DOS.

 

 

GPIB Device Template

The ML2430A Series default primary address is 13. Separate device templates

 

for the primary address of each device can usually be set up separately. The set-

 

tings for the device template for the ML2430A Series are:

 

Terminate read on EOS

 

NO

 

 

 

 

 

 

 

Set EOI with EOS on write

 

YES

 

 

 

 

 

Type of compare on EOS

 

8 bit

 

 

 

 

 

EOS byte

 

0x0A (10 decimal)

 

 

 

 

 

Send EOI at end of write

 

YES

 

 

 

 

 

Readdressing

 

YES

 

 

 

 

 

Secondary address

 

NONE

 

 

 

 

GPIB Card Settings The recommended GPIB card settings for use with the ML2430A Series are:

Terminate read on EOS

NO

 

 

Set EOI with EOS on writes

YES

 

 

Type of compare on EOS

8 bit

 

 

EOS byte

0x0A (10 decimal)

 

 

Send EOI at end of write

YES

 

 

System controller

YES

 

 

Assert REN when SC

YES

 

 

Enable Auto Serial polling

NO

 

 

NI card. Cable length for HS488

OFF

 

 

ML2430A OM

6-5

 

 

Page 92
Image 92
Anritsu ML2430A operation manual Gpib PC Card Setup, 5GPIB PC Card, Gpib Device Template