Anritsu ML2430A operation manual Duty cycle, Rng Hold, Rgh

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 40
Image 40

SENSOR MENU

OPERATION

 

TABLE

Select the offset table number (1-5) when Offset TYPE is set to Ta-

 

[OFFTBL

ble. When a table is selected, additional soft keys become

 

OFFTBR

available:

 

OFFTBU

 

 

OFFVAL]

 

 

 

EDIT

 

 

This will bring up all of the selected offset table’s entries, with their

 

 

associated frequencies and offsets. Select an entry and enter the

 

 

frequency and offset using the keypad.

 

 

CLEAR [OFFCLR]

 

 

When an offset table is selected, CLEAR will set all of the table’s

 

 

elements to zero.

Duty cycle

Applies a duty cycle to the selected sensor. An offset will be applied based on the

 

entered value.

 

SENSOR

Model ML2438A (dual channel) only. Used to select the sensor to

 

 

be configured. Toggles between sensor A and sensor B for all sub-

 

 

menu functions.

 

STATE

ON or OFF

 

[DUTYS]

 

 

DUTY

Delete, Enter, or Cancel. An offset will be applied based on the en-

 

[DUTYS]

tered value. For example, specifying a duty cycle of 50% will alter

 

 

the displayed readings by approximately +3.01 dB.

Rng Hold

[RGH]

NOTE

Rng Hold is not available when SystemSetupmode is set to Source Sweep. In this mode, AUTO ranging is

This function will toggle the sensor between holding the present operating range and Auto Ranging. Auto Ranging automatically selects the best range to take the measurement.

If either sensor is auto ranging, this key will force both sensors to hold their pres- ent operating ranges. If either sensor is held within an operation range, this key will force both sensors to Auto Range.