ML24XXA NATIVE COMMANDS

GPIB OPERATION

val:

Units

Min

Max

 

 

 

 

 

 

dBm

–99.99

+99.99

 

 

 

 

 

dBmV

–53.00

147.00

 

 

 

 

 

dBV

7.00

207.00

 

 

 

 

 

Watts

0.0

50.0

 

 

 

 

Remarks: Sets the high limit. The HLIMS command turns the limits on and off. It is not necessary to enter the units as the limit value is checked against the displayed value. Therefore, if the limits have been set for –10 dBm (HLIM 1, –10) and the display units are subsequently changed from dBm to Watts, the system will still check for the reading to rise above –10, even though the display units type has been changed.

Example: The high limit is set to –10dBm and turned ON. The display is in dBm. A reading of –9.500dBm would pass. If the display is subsequently changed to Watts, a reading of 112.2￿W would fail, because the DISPLAYED value is higher than –10. Limit checking only uses the displayed value and does not change its value even though the display units have changed.

Related

Commands: HLIMS

Query: HLIM? <c>

Returned

String: HLIM <c>,<val>

HLIMS

Turn on/off High limits

CHANNEL

Syntax: HLIMS <c>,<state>

c:1 or 2

state: ON or OFF

Remarks: The HLIMS command turns the limits on and off.

Related

Commands: HLIM

Query: HLIMS? <c>

Returned

String: HLIMS <c>,<state>

HOLD

Graph hold

CHANNEL

6-50

ML2430A OM

 

 

Page 137
Image 137