Anritsu ML2430A Val To 1.0 seconds, Time the system waits after a trigger event has happened

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 166
Image 166

GPIB OPERATION

ML24XXA NATIVE COMMANDS

c:1, 2 or 1&2

val:

0.0 to 1.0 seconds

Remarks:

The time the system waits after a trigger event has happened

 

before taking measurements when in READOUT or POWER vs. TIME

 

mode.

 

Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger

 

together on the same conditions without having to set up two sets of trig-

 

ger data.

Query:

TRGDLY? <c>

Returned

 

String:

TRGDLY <c>,<val>

 

The TRG type commands return the trigger state of the selected channel if

 

the channel is ON. This depends on the settings of the “link triggers” flag

 

and the current mode of the sensor on the selected channel. If the channel

 

is OFF, the stored trigger state of the channel is returned.

TRGGW Set trigger gate width

TRIGGER

Syntax: TRGGW <c>,<val>[units]

c:1, 2 or 1&2

val: 100 ns to 7.0 seconds

Remarks: The length of time the system uses to collect data when in READOUT or POWER vs. TIME mode. The default value is 20 ms.

Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger together on the same conditions without having to set up two sets of trig- ger data.

Query: TRGGW? <c>

Returned

String: TRGGW <c>,<val>

The TRG type commands return the trigger state of the selected channel if the channel is ON. This depends on the settings of the “link triggers” flag and the current mode of the sensor on the selected channel. If the channel is OFF, the stored trigger state of the channel is returned.

TRGLVL

Set trigger level

TRIGGER

Syntax: TRGLVL <c>,<val>

c:1, 2 or 1&2

val: –30 to +20 dBm

ML2430A OM

6-79

 

 

Page 166
Image 166
Anritsu ML2430A Val To 1.0 seconds, Time the system waits after a trigger event has happened, Ger data, Trglvl