GPIB OPERATION

ML24XXA NATIVE COMMANDS

 

 

MANUAL (manual push button trigger)

 

 

 

CONT (continuous)

 

 

Remarks:

This command is overridden by the TR0, TR1 and TR2 commands when

 

 

in READOUT or POWER vs. TIME mode. If TR3 is sent, the trigger source

 

 

reverts back to the previously selected type of triggering.

 

 

 

Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger

 

 

together on the same conditions without having to set up two sets of trig-

 

 

ger data.

 

 

Query:

TRGSRC? <c>

 

 

Returned

 

 

 

String:

TRGSRC <c>,<source>

 

 

 

The TRG type commands return the trigger state of the selected channel if

 

 

the channel is ON. This depends on the settings of the “link triggers” flag

 

 

and the current mode of the sensor on the selected channel. If the channel

 

 

is OFF, the stored trigger state of the channel is returned.

 

TRGTYP

Set Trigger type

 

 

 

TRIGGER

 

Syntax:

TRGTYP <c>,<type>

 

 

c:

1, 2 or 1&2

 

 

type:

RISE

 

 

 

FALL

 

 

Remarks:

Sets the control type of the trigger used when the source is set to either

 

 

INTA or INTB (internal A or B) in READOUT or POWER vs. TIME mode.

 

 

Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger

 

 

together on the same conditions without having to set up two sets of trig-

 

 

ger data.

 

 

Query:

TRGTYP? <c>

 

 

Returned

 

 

 

String:

TRGTYP <c>,<type>

 

 

 

The TRG type commands return the trigger state of the selected channel if

 

 

the channel is ON. This depends on the settings of the “link triggers” flag

 

 

and the current mode of the sensor on the selected channel. If the channel

 

 

is OFF, the stored trigger state of the channel is returned.

 

TRGXTTL

Set external trigger edge type

 

TRIGGER

 

Syntax:

TRGXTTL <c>,<type>

 

 