GPIB OPERATION

ML24XXA NATIVE COMMANDS

range 3 upper, range 3 lower, range 4 upper, range 4 lower, range 5 up- per, range 5 lower.

REL

Relative control

 

 

CHANNEL

 

Syntax:

REL <c>,<mode>

 

 

 

c:

1 or 2

 

 

 

mode:

0 Turn OFF

 

 

 

 

1 Turn ON and reference

 

 

 

 

2 Turn ON, use old references if not first time.

 

 

 

Remarks:

Turns relative ON or OFF, or references the zero point.

 

 

 

 

REL1 and REL2 toggle between relative and absolute measurements.

 

 

Sending the RELx command when in dB mode will make the meter apply

 

 

the last used RELATIVE value. This relative value is used thereafter until it

 

 

is replaced by another one in the same manner. This allows the user to re-

 

 

fer to a previously referenced value, without the meter resetting itself back

 

 

to a 0.00 display.

 

 

 

Query:

REL? <c>

 

 

 

Returned

 

 

 

 

String:

REL <c>,<mode>

 

 

RFCAL

Turn RF reference calibrator ON or OFF

 

CALIBRATION

 

Syntax:

RFCAL <state>

 

 

 

state:

ON or OFF

 

 

 

Remarks:

Turns on or off the RF reference calibrator.

 

 

Query: RFCAL?

Returned

String: RFCAL <state>

RGH

Range Hold Sensor

SENSOR

Syntax: RGH <s>[,<val>]

s:A or B

val: 0 to 5 (0 to 6 for universal power sensor only) (0 = AUTO)

ML2430A OM

6-67

 

 

Page 154
Image 154