Anritsu ML2430A operation manual Sta, Wat, Zaj

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 181
Image 181

ML4803A EMULATION COMMANDS

GPIB OPERATION

Remarks: When SRQ0 is issued, the SRQ will no long turn off and on with each reading. The SRQ is set back on by the SRQ1 command or by requesting data.

STA

Restart averaging reading.

WAT

Sets the display channel units to Watts.

Remarks: Turns off relative mode. Relative is not available in this mode.

ZAJ

Zero the sensor.

Remarks: During the zero operation, the zero bit in the status byte is set. When the zero operation is completed, the zero bit in the status byte is cleared.

When emulating the ML4803, the ML2430 may take longer to zero a sen- sor than the ML4803 itself. When performing a zero, the status byte should be used to identify when zeroing is complete.

6-94

ML2430A OM

 

 

Page 181
Image 181
Anritsu ML2430A operation manual Sta, Wat, Zaj