GPIB OPERATION

HP 436A EMULATION COMMANDS

I

Trigger without settling.

Remarks: Triggers a single reading which is added to the internal digital filter and the updated filter power level is returned on the GPIB. After an I command, the instrument returns to standby mode (H).

OI Identification

Remarks: Ask for identification of current operating mode. Responds with “HP436.”

R

Free run mode

Remarks: Sets the ML2430A Series back into free run mode to continuously take measurements and output data.

T

Trigger with settling

Remarks: Triggers a new series of readings; enough to update the internal digital fil- ter for a noise free reading at the current power level. The value is then re- turned on the GPIB and returns to standby mode (H).

V

Free run mode with settling

Remarks: Sets the ML2430A Series back into free run mode to continuously take measurements and output data after running a settling routine.

Z

Zero sensor

Remarks: Zero out the noise from the sensor. When zeroing the ML2430 in HP 436 emulation mode, the 'Z1T' sequence followed by the '9+DI' described in the HP 436 manual must be followed.