GPIB OPERATION | HP 438A EMULATION COMMANDS |
Remarks: Triggers a single reading which is added to the internal digital filter and the updated filter power level is returned on the GPIB. After a TR1 command, the instrument returns to TR0 standby mode.
Related TR0, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
Commands:
TR2 | Trigger with a settling delay |
Syntax: TR2
Remarks: Triggers a new series of readings; enough to update the internal digital fil- ter for a noise free reading at the current power level. The value is then re- turned on the GPIB and returns to TR0 standby mode.
Related
Commands: TR0, TR1, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR3 | Trigger free run |
Syntax: TR3
Remarks: Sets the ML2430A Series back into free run mode on both channels.
Related
Commands: TR0, TR1, TR2, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
ZE | Zero sensors |
Syntax: ZE
Remarks: Zero all connected sensors. The ML2430, when emulating the HP 438, may take longer to Zero a sensor than the HP 438 itself. When performing a zero the status byte should be used to identify when ZEROing is com- plete.
ML2430A OM |
|
|
|