HP 437B EMULATION COMMANDS

GPIB OPERATION

Remarks: Same as the *CLS command. Resets all of the GPIB status registers and clears the input queue.

CT

Clear the cal factor table

Syntax: CT<table_number>

table number: 0 to 9

Remarks: Clears the specified cal factor table to a single 50mhz entry at 100%. Since the ML2430A stores the cal factor table information in the sensor, this data must be saved to the sensor by using the ‘EX’ command or an additional command ‘SV’, or the data could be lost. The saving of the cal factor table data to the sensor can be done at the end of all updates to a particular table.

DA

Display All

Syntax: DA

Remarks: Turns on all the segments of the display to verify proper operation. The dis- play is returned to normal when another command is sent.

DC

Duty Cycle state

Syntax: DC<state>

state: 0 = OFF 1 = ON

Remarks: Turns on or off application of the duty cycle to the sensor data.

DD

Display disable

Syntax: DD

Remarks: Turns the display off to allow faster measurements to be taken.

Related

Commands: DE, DF

6-104