GPIB OPERATION

ML24XXA NATIVE COMMANDS

–10.234 becomes 10479, converted to hexadecimal FFFFD711. See the programming examples for more detail.

OFFTYP

Offset type to use

SENSOR

Syntax: OFFTYP <s>,<type>

s:A or B

type: OFF FIXED TABLE

Remarks: Selects the type of offset to use.

OFF = No offset to be used.

FIXED = Use the fixed value (OFFFIX) specified.

TABLE = Use the Offset table (OFFTBL) specified.

Query: OFFTYP? <s>

Returned

String: OFFTYP <s>,<type>

OFFVAL

Sensor Offset Value

SENSOR

Syntax: OFFVAL <s>

s:A or B

Remarks: Returns the Offset value being applied to the specified sensor.

Related

Commands: OFFTBL, OFFTYP

OGBD

Output Graph Binary Data

DATA OUTPUT

Syntax: OGBD

Remarks: Output the next complete set of graph data to the GPIB in the long integer form of 1024 bits per dB as a definite length arbitrary block response data. The C programming example 'Binary output decoding' on page 6-133shows how to extract the binary data. The response form is as follows : OGBD <#><length><number_of_bytes><data_byte_1><data_byte_2>

...<data_byte_n><\n>

<length> number of ASCII characters that make up the number_of_bytes value

<number_of_bytes> number of bytes of data contained in rest of the string <data_byte_n> four of these values makes up the long integer.

ML2430A OM