ML24XXA NATIVE COMMANDS | GPIB OPERATION |
Syntax: GTLVL <val>
val:
Remarks: When the system trigger in profile mode is set to either INTA or INTB (in- ternal sensor A or B) it will trigger on a power level given by the sensor. This command sets the level.
Related
Commands: GTSRC, GTTYP
Query: GTLVL?
Returned
String: GTLVL <val>
GTSRC | Set Profile Trigger source | TRIGGER |
Syntax: GTSRC <source>
source: INTA INTB EXTTTL MANUAL CONT
Remarks: INTA = internal sensor A
INTB = internal sensor B
EXTTTL = external BNC TTL trigger input
MANUAL = manual push button trigger
CONT = continuous
MANUAL trigger only functions correctly on
The display shows an ‘x’ marking the trigger point. This trigger point mark rotates as the profile data is updated, changing between 'x' and '+' on each data update. On rapid updates, the trigger point mark may appear like a star (*), as it is rotating so quickly. In manual, external or GPIB triggered displays, the mark rotates at a slower rate and each true data update can be seen.
The GTSRC setting is overridden by the Group Execute Trigger GPIB com- mon command (GET), *TRG, TR0, TR1 and TR2 commands. The TR3 command will return the system to its previous state if the TR0 (Trigger hold) command has been used.
Query: GTSRC?
Returned
String: GTSRC <source>
ML2430A OM | |
|
|