Chapter 5 Procedures
INTRODUCTION | This chapter presents some common procedures for use with the |
| ML2430A Series Power Meter. These procedures refer to the ML2430A |
| Series front and rear panel connectors and front panel keys and menus |
| as explained in Chapter 3, Connections, and Chapter 4, Front Panel Op- |
| eration. The operator should be familiar with the front and rear panel lay- |
| outs and with the use of the keys and menus before |
| attempting these procedures. |
POWER | To perform a power measurement, follow these steps: |
MEASUREMENT | Connect the sensor(s) as described in Chapter 3, Connections. |
| |
| Configure the meter for the application. Refer to Chapter 4, Front |
| Panel Operation, for specific configuration options. The simplest op- |
| eration is obtained with SENSORSETUPMODE set to DEFAULT. |
| Power readings are continuous with the default setting. |
| Zero the sensor(s) as described in Section |
| Calibrate the sensor(s) as described in Section |
| Measure power. |
ZEROING THE | Zero the sensor before making power measurements, particularly when |
SENSOR | operating within the lower 20 dB dynamic range of the power sensor. If |
| frequent low level measurements are being made, it is advisable to check |
| the sensor zeroing often and repeat as necessary. |
| To zero the sensor, connect it to the UUT (Unit Under Test) test port, and |
| remove RF power from the connection to a level 20 dB below the tangen- |
| tial noise floor of the power sensor. For |
| sensors, this level is less than |
| It is preferable to leave the sensor connected to the UUT test port so that |
| ground noise and thermal EMF are zeroed out of the measurement. Alter- |
| nately, in order of preference, the sensor can be connected to: |
| A grounded connector on the UUT, |
| the ML2430A Series Calibrator connector, |
| disconnected from any signal source. |
| When a new sensor is attached, the message SENSOR x NOT ZEROED |
| (where x = A or B as appropriate) is displayed. If a sensor is removed and |
| then reconnected, the message is not displayed. |
| The sensor can either be zeroed, or zeroed and calibrated in the same |
| operation. |
ML2430A OM | |
|
|