ML24XXA NATIVE COMMANDS

GPIB OPERATION

Syntax: MODRED <count>

count: 0 to 10

Remarks: Sets the number of retrys after a failure to connect. The delay between retrys is set using MODDEL. See Section 5-10 for more information on mo- dem operation.

Query: MODRED?

Returned

String: MODRED <count>

MODRNG Autodial enable for range error

SYSTEM

Syntax: MODRNG <state>

value: TRUE or FALSE

Remarks: When set to TRUE, produces an SRQ and autodials the phone number (set with MODPH) when a sensor range error occurs. See Section 5-10 for more information on modem operation.

Query: MODRNG?

Returned

String: MODRNG <state>

MXGDB

Output Max Graph Binary Data

DATA OUTPUT

Syntax: MXGDB

Remarks: Available in graph modes only. Outputs in binary form the max graph data to the GPIB in the long integer form of 1024 bits per dB as a definite length arbitrary block response data. The C programming example 'Binary output decoding' on page 6-133shows how to extract the binary data. The re- sponse form is as follows :

MXGDB <#><length><number_of_bytes><data_byte_1><data_byte_2>

...<data_byte_n><\n>

<length> number of ASCII characters that make up the number_of_bytes value

<number_of_bytes> number of bytes of data contained in rest of the string <data_byte_n> four of these values make up the long integer.

For example: FF FF D1 64 = –11932 As it is based on 1024 per dB, divide by 1024 to get the dB value (–11.652).

MXGD

Output Max Graph Data

DATA OUTPUT

6-56

ML2430A OM

 

 