GPIB OPERATION

ML4803A EMULATION COMMANDS

 

 

When selecting GPIB emulation modes, the instrument configures itself to

 

 

the preset conditions of the instrument to be emulated. For example, when

 

 

selecting HP 438A emulation, the front panel menus pass through the pre-

 

 

sets for the HP 437B (which presets sensor A to dBm) then selects HP

 

 

438A emulation (which presets sensor A to Watts).

 

 

NOTE

 

 

This command must be entered using the 488.2 format;

 

 

that is, EMUL<ws><mode> (<ws> = white space).

MCA

Set the cal factor value at the specified memory location in

 

dBm.

 

 

Syntax:

MCA<mem><;><value>

 

mem:

Memory location 1 to 30.

 

value:

Cal factor value in dBm.

 

Remarks:

Set the cal factor value at memory store address <mem> to <value> dBm.

MCC

Clears the cal factor value at the specified memory loca-

 

tion.

 

 

Syntax:

MCC<mem>

 

mem:

Memory location 1 to 30.

 

Remarks:

Clears the cal factor value at memory store <mem> to 0.0 dBm.

MCO

Clears the offset value at the specified memory location.

 

Syntax:

MCO<mem>

 

mem:

Memory location 1 to 30.

 

Remarks:

Clears the offset value at memory store <mem> to 0.0 dBm.

MCQ

Clears the frequency value at the specified memory loca-

 

tion.

 

 

Syntax:

MCQ<mem>

 

mem:

Memory location 1 to 30.

 

Remarks:

Clears the frequency value at memory store <mem> to 0.1MHz.

ML2430A OM

6-87

 

 

Page 174