GPIB OPERATION | HP 437B EMULATION COMMANDS |
Remarks: Stores the present configuration to the selected register.
Example: To store the current instrument configuration in register 2:
ST2EN
Related
Commands: RC
SV | Save cal factor table |
Syntax: SV
Remarks: Since the ML2430A stores the cal factor tables in the sensors, this command forces the edits to a cal factor table to be saved to the sensor. The operation can take a couple of seconds to complete.
TR0 | Trigger hold mode |
Syntax: TR0
Remarks: Sets both channels to trigger hold mode. It does not trigger until it receives a TR1 or TR2 or GET (group executive trigger), *TRG or TR3 command. If it receives a TR3 command it reverts back to the trigger mode it was in be- fore the TR0 command was sent. If the REM line is low, this command has no effect.
Related
Commands: TR1, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR1 | Trigger immediate |
Syntax: TR1
Remarks: Triggers a single reading which is added to the internal digital filter and the updated filter power level is returned on the GPIB. After a TR1 command, the instrument returns to TR0 standby mode.
Related
Commands: TR0, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR2 | Trigger with a settling delay |
Syntax: TR2
ML2430A OM | |
|
|