ML24XXA NATIVE COMMANDS

GPIB OPERATION

port: 1 or 2

c:1 or 2

Remarks: This command changes the channel represented by BNC output modes that can take data from either channel 1 or 2, such as “Analog Output” and “Pass/Fail” modes.

Query: OBCH? <port>

Returned

String: OBCH <port>,<c>

OBDSP

BNC analog output display stop value

BNC

Syntax: OBDSP <port>,<units>,<val>

port: 1 or 2 units: W (Watts)

DB (dB)

DBM (dB) DBUV (dB￿V) DBMV (dBmV)

val: 0 to 50W –70 to 47dB –23 to 94 dBmV 37 to 154 dB￿V

Remarks: Sets up the stop value for the analog out scale of the display.

Query: OBDSP? <port>

Returned

String: OBDSP <port>,<units>,<val>

OBDST

BNC analog out display start value

BNC

Syntax: OBDST <port>,<units>,<val>

port: 1 or 2 units: W (Watts)

DB (dB)

DBM (dB) DBUV (dB￿V) DBMV (dBmV)

val: 0 to 50W –70 to 47dB –23 to 94 dBmV 37 to 154 dB￿V

Remarks: Sets up the start value for the analog out scale of the display.