Anritsu ML2430A Syntax FA, Syntax FH, Related Commands FM, FH, FA, Related Commands FH, FA

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 194
Image 194

GPIB OPERATION

HP 437B EMULATION COMMANDS

FA

Auto average

Syntax: FA

Remarks: Automatic Filter on. Allows the system to automatically select the filter used to reduce the jitter in the display.

Related

Commands: FM, FH

FH

average hold

Syntax: FH

Remarks: Hold filter sets the filter mode to Manual from Auto, but retains the auto fil-

ter setting. This function is the same as the AVGM command.

Related

Commands: FM, FH, FA

FM

Set average value

Syntax: FM<val>EN

val: 1 to 512

Remarks: Sets the filter length for the averaging of sensor data. For HP 437B emula- tion, the command accepts 1 to 512 in 2-to-the-power steps. For example, 1, 2, 4, 8, 16,...256, 512.

Related

Commands: FH, FA

FR

Frequency of the input signal

Syntax: FR<val><units>

val:

units: GZ (GHz) MZ (MHz) KZ (KHz) HZ (Hz)

EN (Hz)

Remarks: Sets the frequency of the input signal so that the correct cal factor is used.

ML2430A OM

6-107

 

 

Page 194
Image 194
Anritsu ML2430A operation manual Syntax FA, Syntax FH, Related Commands FM, FH, FA, Related Commands FH, FA