GPIB OPERATION

ML24XXA NATIVE COMMANDS

Syntax: ON <c>,<val>

c:1 or 2

val: 1 to 1000

Remarks: Readout and Power vs. Time modes only. This command returns the speci- fied number of readings for the specified channel. The readings are first assembled, and then passed to the GPIB as a whole, with a line feed char- acter (hex 0x0a) marking the end of the string.

Example: ON 1, 9

This example will return:

–10.234, –10.234, –10.235, –10.238, -10.250, –10.270, –10.500, –10.934, –12.234<0x0a>

OPMD

ML2430A Series operation mode

SYSTEM

Syntax: OPMD <mode>

mode: DIGIT PROFILE PWRTIM SRCSWP

Remarks: This command selects the ML2430A Series operation mode (data collec- tion method).

DIGIT = digital read out of channel data PROFILE = profile of graphic display PWRTIM = graph of channel power versus time SRCSWP = source sweep graphic display

To use Graph output commands, the ML2430A Series must be in Profile or Power vs. Time mode. To use the output channel data commands, the ML2430A Series must be in the digital readout mode.

Related

Commands: GRMD

Query: OPMD?

Returned

String: OPMD <mode>

PRINT

Send details to the connected printer.

SYSTEM

Syntax: PRINT

Remarks: The type of printout depends on the operation mode currently selected. In all modes, the printout includes a header with the current sensor settings

ML2430A OM

6-65

 

 

Page 152
Image 152
Anritsu ML2430A Example on 1, Opmd, Mode Digit Profile Pwrtim Srcswp, Related Commands Grmd Query OPMD?, Syntax Print