Marathon Monitors Inc.
AACC 2000 Carbon Nov. 1, 1997
83
Name Description Values Meaning
CALCAL Calibration
In this mode you can
1. Calibrate the instrument using a mV source - rcAL or ref source cal.
2. Offset the calibration to account for errors in actual sensor
measurement and a ref sensor - UCAL or user calibration
3. Return to factory set calibration - FACT or factory set calibration.
rcAL Calibration
point nonE No calibration
PV Calibrate main Process Value input.
PV.2 Calibrate DC input, or PV 2.
1A.Hi Calibrate DC output high - Module 1
1A.Lo Calibrate DC output low - Module 1
2A.Hi Calibrate DC output high - Module 2
2A.Lo Calibrate DC output low - Module 2
3A.Hi Calibrate DC output high - Module 3
3A.Lo Calibrate DC output low - Module 3

INPUT CALIBRATION

ForCAL=PV’, or ‘PV.2’, the following parameters apply.
PV PV Calibration Value IdLE Idle
mv.L Select 0mV as the calibration
point
mv.H Select 50mV as the calibration
point
V 0Select 0Volt as the calibration
point
1. Select calibration value V 10 Select 10V as the calibration
point
2. Apply specified input CJC Select 0oC CJC calibration point
3. Press to step to GOrtd Select 400 as the calibration
point
HI 0High impedance: 0Volt cal’n
point
HI 1.0 High impedance: 1.0 Volt cal’n
point
See Note below. FACt Restore factory calibration
GO Start calibration no Waiting to calibrate PV point
Goto User
calibration
table-See also
chapter 7
Go to input
Calibation table
Go to
DC Output
Calibration
table