Offset calibration
Offset calibration is used to apply a single fixed offset over the full display range of the controller.
Displayed Value
Factory Calibration
Fixed Offset
Input
To calibrate, proceed as follows:
1.Connect the input of the controller to the source device to which you wish to calibrate.
2.Set the source to the desired calibration value.
3.The controller will display the current measurement of the value.
4.If the displayed value is correct, then the controller is correctly calibrated and no further action is necessary. If it is incorrect, then follow the steps shown below.
Select ‘FuLL’ access level, as described in Chapter 3.
Input list header
Press until you reach the input list header.
x 3 | Press | until you reach the ‘ | CAL | ’ display. | ||
|
| |||||
| Calibration type |
|
|
| ||
| ∙ | FACt: |
| Factory Calibration | ||
| ∙ | USEr: |
| User Calibration |
| |
| Use |
| or | to select ‘FACt’. | ||
| Selecting ‘FACt’ reinstates the factory | |||||
| calibration and allows the application of a single | |||||
| fixed offset. |
|
|
| ||
| Press | continued on the next page |
|