Marathon Monitors Inc.
User Calibration Enable
The User calibration facility must first be enabled in configuration level by setting the parameter ‘UCAL' in the input conf list to 'YES'. This will make the User calibration parameters visible in Operator ‘FuLL’ level.
Select configuration level as shown in Chapter 5, Configuration.
The Calibration Configuration List
Press 
 until you reach the 
Press 
 until you reach ‘UCAL’.
User Calibration Enable
Use 
 or 
 to select:
∙YES: Calibration enable
∙no: Calibration disabled
  | Press  | and  | together to go to the Exit  | 
+  | display.  | 
  | 
  | 
Exit configuration
Use 
 or 
 to select ‘YES’ to return to Operator level.
AACC 2000 Carbon  | 86  | Nov. 1, 1997  |