Marathon Monitors Inc.

Calibration points and Calibration offsets

If you wish to see the points at which the User calibration was performed and the

value of the offsets introduced, then these are shown in Configuration, in ‘CAL- Conf’.

The parameters are:

Name

Parameter description

Meaning

pt1.L

Low calibration point for Input 1

The factory calibration point at which the

 

 

low point offset was performed.

pt1.H

High calibration point for Input 1

The factory calibration point at which the

 

 

high point offset was performed.

OF1.L

Offset Low for Input 1

Calculated offset, in display units.

OF1.H

Offset High for Input 1

Calculated offset, in display units.

pt2.L

Low calibration point for Input 2

The factory calibration point at which the

 

 

low point offset was performed.

pt2.H

High calibration point for Input 2

The factory calibration point at which the

 

 

high point offset was performed.

OF2.L

Offset Low for Input 2

Calculated offset, in display units.

OF2.H

Offset High for Input 2

Calculated offset, in display units.

Note: The value of each of the parameters in the above table may also be altered by using the / buttons.

AACC 2000 Carbon

92

Nov. 1, 1997

Page 92
Image 92