See table on the right for additional parameters.

Set Offset 1

Use or to set the offset value of Process Value 1 (PV1).

The offset value is in display units Press

Set Offset 2

Use or to set the offset value of Process Value 2 (PV2), if configured.

The offset value is in display units.

Press

The table below shows the parameters which appear after ‘OFS.2’. These are all read only

values and are for information. Press

to step

through them.

 

 

 

 

mV.1

IP1 measured value (at terminals)

 

mV.2

IP2 measured value (at terminals), if DC

 

input in Module 3 position

 

CJC.1

IP1 Cold Junction Compensation

 

CJC.2

IP2 Cold Junction Compensation

 

Li.1

IP1 Linearised Value

 

Li.2

IP2 Linearised Value

 

PV.SL

Shows the currently selected input

 

 

 

If you do not want to look at these parameters,

then press

and this returns you to the ‘iP-

LiSt’ header.

To protect the calibration against unauthorised adjustment, return to Operator level and make sure that the calibration parameters are hidden. Parameters are hidden using the ‘Edit’ facility described in Chapter 3, Access Levels

AACC 2000 Carbon