Technical Specification

Environmental ratings

 

 

Panel sealing:

Instruments are intended to be panel mounted. The

 

rating of panel sealing is IP65, (EN 60529), or 4X, (NEMA

 

250).

 

Operating temperature:

0 to 55oC. Ensure the enclosure provides adequate

ventilation.

 

 

Relative humidity:

5 to 95%, non condensing.

Atmosphere:

The instrument is not suitable for use above 2000m

 

or in explosive or corrosive atmospheres.

Equipment ratings

 

 

Supply voltage:

100 to 240Vac -15%, +10%, or optionally:

Supply frequency:

48 to 62Hz.

Power consumption:

15 Watts maximum.

Relay 2-pin (isolated):

Maximum: 264Vac, 2A resistive. Minimum: 12Vdc, 100mA.

Relay changeover (isolated): Maximum: 264Vac, 2A resistive. Minimum: 6Vdc, 1mA.

Triac outputs (isolated):

30 to 264Vac. Maximum current: 1A resistive.

Leakage current:

The leakage current through triac and relay contact

 

suppression components is less than 2mA at 264Vac, 50Hz.

Over current protection:

External over current protection devices are required that

 

match the wiring of the installation. A minimum of 0.5mm2 or

 

16awg wire is recommended. Use independent fuses for the

 

instrument supply and each relay or triac output. Suitable

 

fuses are T type, (EN 60127 time-lag type) as follows;

 

Instrument supply: 85 to 264Vac, 2A, (T).

 

Relay outputs: 2A (T). Triac outputs: 1A (T).

Low level i/o:

All input and output connections other than triac and relay are

 

intended for low level signals less than 42V.

Single logic output:

18V at 24mA. (Non-isolated.)

DC output (Isolated):

0 to 20mA (600Ω max), 0 to 10V (500Ω min).

DC output (Non isolated):

0 to 20mA (600Ω max), 0 to 10V (500Ω min).

Fixed digital inputs:

Contact closure. (Non isolated.)

Triple contact input:

Contact closure. (Isolated.)

Triple logic input:

11 to 30Vdc. (Isolated.)

DC or 2nd PV input:

As main input plus 0-1.6Vdc, Impedance, >100MΩ.

 

(Isolated.)

Potentiometer input:

0.5V excitation, 100Ω to 1.5kΩ Potentiometer. (Isolated.)

Transmitter supply:

24Vdc at 20mA. (isolated.)

Strain gauge supply:

10Vdc. Minimum bridge resistance 300Ω. (Isolated.)

Digital Communications:

EIA-232, 2-wire EIA-485 or 4-wire EIA-485 (All isolated).