AppendixA Specificationsfor AT-MIO-16XE-10and AT-AI-16XE-10
©NationalInstruments Corporation A-23 ATE Series User Manual
CMRR, DC to 60 Hz
Dynamic Characteristics
Bandwidth
Allg ains.......................................... 255 kHz
Settling time for full-scale step
System noise (LSBrms including quantization noise)
Range CMRR (Bipolar) CMRR(Unipolar)
20 V 92 dB
10 V 97 dB 92dB
5V 97 dB
4V 101 dB
2V 104 dB 101 dB
1V 105 dB 104 dB
100 mV to 500 mV 105 dB 105 dB
Accuracy*
±0.00076%
(±0.5 LSB)
±0.0015%
(±1 LSB)
±0.0016%
(±4LSB)
40µs max 20 µs max 10 µs max
*Accuracy values valid for source impedances < 1 k. Referto the Multiple-Channel
ScanningConsiderations section of Chapter 3, Hardware Overview,formore
information.
Range Bipolar Unipolar
2to20V 0.6 0.8
1V 0.7 0.8
400 to 500 mV 1.1 1.1
200 mV 2.0 2.0
100 mV 3.8