Chapter4 ConnectingSignals
ATE Series User Manual 4-38 ni.com
CONVERT* SignalAny PFI pin can externally input the CONVERT* signal, which is
available as an output on the PFI2/CONVERT* pin.
Refert o Figures 4-13 and 4-14 for the relationship of CONVERT*to the
DAQsequence.
As aninpu t, theC ONVERT*signal is configured in the edge-detection
mode. Youcan select any PFI pin as the source for CONVERT* and
configurethe polarity selection for either ri sing orfalling edg e. The
selectededge of the CONV ERT*signal initiates an A/D conversion.
Asan output, the CONVERT* signal reflects the actual convert pulse that
is connected to the ADC, evenif the conversions are being externally
generated byanother PFI. Th eo utput is an activelow pulse with a pulse
width of5 0 to100 ns. This outputis s etto high-impedance at startup.
Figures 4-21 and 4-22 showthe input and outp ut timing requirements for
the CONVERT*signal.
Figure4-21. CONVERT* Input Signal Timing
Figure4-22. CONVERT* Output Signal Timing
Rising-Edge
Polarity
Falling-Edge
Polarity
tw=10 ns minimum
tw
tw=50to150ns
tw