Chapter4 ConnectingSignals
©NationalInstruments Corporation 4-19 ATE Series User Manual
Figure4-5. Summary of Analog Input Connections
+
+
V1
ACH
AISENSE
AIGND
+
+
V1
ACH
AISENSE
AIGND
R
Seetext for information on bias resistors.
+
+
V1
ACH
AIGND
+
+
V1
ACH
+V
g
AIGND
Ground-looplosses, Vg, are added to
measuredsignal.
NOTRECOMMENDED
+
+
V1
ACH(+)
ACH()
AIGND
+
+
V1
ACH(+)
ACH()
AIGND
R
Seetext for information on bias resistors.
SignalSource Type
FloatingSignal Source
(NotConnected to Building Ground)
GroundedSignal Source
Examples
UngroundedThermocouples
SignalConditioning with
IsolatedOutputs
BatteryDevices
Examples
Plug-inInstruments with
NonisolatedOutputs
Input
Differential
(DIFF)
Single-Ended
Ground
Referenced
(RSE)
Single-Ended
Nonreferenced
(NRSE)