Chapter4 ConnectingSignals
ATE Series User Manual 4-26 ni.com
Figure 4-9 showshow to connect a grounded signal source to an
ATE Series device channel configured for NRSE mode.
Figure4-9. Single-EndedInput Connections for Ground-Referenced Signal
Common-Mode Signal Rejection Considerations
Figures 4-6 and 4-9 show connections for signal sources that are already
referenced to some ground point with respect to the AT E Series device.
In thesecases, the PGIA can reject any voltage caused by ground potential
differencesbetween the signal source and the device. In addition, with
differentialinput c onnections,the PGIA canreject common-mode noise
pickup in the leads connecting the signal sources to the device. The PGIA
can reject common-mode signals as long as Vin+ and Vinare both within
±11 V of AIGND. The AT-MIO-16XE-50 has the additional restriction