Differential Connections for Ground-ReferencedSignal Sources

Figure 4-6 shows how to connect a ground-referenced signal source to

an ATE Series device channel configured in DIFF input mode.

Figure4-6. Differential Input Connections for Ground-Referenced Signals

Withthis type of connection, the PGIA rejects both the common-m ode

noise in the signal and the ground potential difference between the signal

sourcean dthe AT E Series device ground, shown as Vcm in Figure 4-6.