Philips M3861A, M3860A manual Preparation

Models: M3860A M3861A

1 102
Download 102 pages 22.88 Kb
Page 22
Image 22

3-2

N O

 

 

 

 

F

O

 

F

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step 1: Preparation

Press the On/Off button to turn on the HeartStart FR2+ Defibrillator. Follow the instructions provided by the FR2+ voice and screen prompts in the order indicated.

Remove clothing from the patient's upper body. Wipe moisture from the patient's skin and clip or shave excessive chest hair, if necessary.

If the patient appears to be under eight years of age or 55 lbs (25 Kg), use M3870A FR2 infant/child reduced-energy defibrillator pads, if available. If the child appears older/larger, use adult defibrillator pads. DO NOT DELAY

TREATMENT TO DETERMINE THE CHILDS EXACT AGE/WEIGHT.

Open the defibrillator pads package. Check to see that the pads and attached cable and connector are undamaged. Pull off the protective backing from the defibrillator pads and check that the gel has not dried out. If the pads are damaged or the gel has dried out, use a new set of pads.

Place each pad on the patient. The pads must be placed with the sticky side on the patient’s skin. I M P O R T A N T : Refer to the drawing on each pad for correct positioning. For adult patients, one pad goes just below the patient's right collarbone, and the other one goes over the patient’s ribs in line with the armpit and below the left breast. For children under eight years old, one pad is centered on the chest between the nipples, and the other on the back between the scapulae (shoulder bones).

Connect the pads to the FR2+. Insert the defibrillator pads connector firmly into the connector socket. A flashing light shows you where the socket is located, at the top left of the FR2+.

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Page 22
Image 22
Philips M3861A, M3860A manual Preparation