category | nominal specifications |
|
|
Controls | On/Off button |
| Shock button |
| Option buttons |
|
|
LED Indicators | Connector socket LED, flashes to indicate socket location. LED is covered when |
| defibrillator pad connector is properly inserted. Shock button LED flashes when |
| defibrillator is armed. |
|
|
Audio Speaker | Provides voice prompts (volume is adjustable via Setup screen). |
|
|
Beeper | Chirps when a selftest has failed. |
| Provides various warning beeps during normal use. |
|
|
Status Indicator | Status indicator LCD displays device readiness for use. |
|
|
Low Battery Detection | Automatic during daily periodic selftesting. |
|
|
Low Battery Indicator | Solid or flashing red X Status Indicator on front panel; screen display LOW |
| BATTERY or REPLACE BATTERY warning, as appropriate. |
|
|
| Accessories Specifications |
| |
| M3863A FR2 standard battery |
| |
|
|
| |
category | nominal specifications |
| |
|
|
| |
Battery Type | 12 VDC, 4.2 Ah, lithium manganese dioxide. Disposable, |
| |
|
|
| |
Capacity | When new, a minimum of 300 shocks or 12 hours of operating time at 77° F |
| |
| (25° C). |
| |
|
|
| |
Shelf Life | Typically, 5 years from date of manufacture when stored under standby | stems | |
(prior to installation) | environmental conditions in original packaging. | ||
| |||
|
|
| |
Standby Life | Typically, 5 years. >4 years when stored under standby environmental conditions | Sy | |
(after installation) | (battery installed, FR2+ unused). | dical | |
|
| ||
Status Indicators | Good battery: flashing black hourglass on the front panel of the FR2+. | ||
Me | |||
| Low battery: flashing red X on the front panel of the FR2+. | ||
| ps | ||
| Dead battery: solid red X on the front panel of the FR2+. | ||
| P hili | ||
|
| ||
Storage Temperature | 32° to 109° F (0° to 43° C). | ||
| |||
|
|
|