B-4

category

nominal specifications

 

 

Disarm (advanced mode)

Once charged, the HeartStart FR2+ will disarm if:

 

in advanced mode ANALYZE

 

• the manual disarm button is pressed, OR

 

• a patient’s heart rhythm changes to non-shockable rhythm, OR

 

• a shock is not delivered within 30 seconds after the FR2+ is armed, OR

 

• the On/Off button is pressed to turn off the FR2+, OR

 

• the defibrillator pads are removed from the patient, OR

 

• the pads connector is disconnected from the FR2+.

 

in advanced mode CHARGE (M3860A only)

 

• the manual disarm button is pressed, OR

 

• a shock is not delivered within 30 s after charging, OR

 

• the On/Off button is pressed to turn off the FR2+, OR

 

• the defibrillator pads are removed from the patient, OR

 

• the pads connector is disconnected from the FR2+.

 

 

Shock Delivery Vector

Via adult defibrillator pads placed in the anterior-anterior (Lead II) position or via

 

FR2 infant/child reduced-energy defibrillator pads placed in the anterior-posterior

 

position.

 

ECG Analysis System

 

category

nominal specifications

 

 

Function

Evaluates impedance of defibrillator pads for proper contact with patient skin, and

 

evaluates the ECG rhythm and signal quality to determine if a shock is appropriate.

 

 

Protocols

Follows pre-programmed settings to match local EMS guidelines or medical

 

protocols. The settings can be modified using the setup options.

 

 

Shockable Rhythms

Ventricular fibrillation (VF) and certain ventricular tachycardias, including

 

ventricular flutter and polymorphic ventricular tachycardia (VT). The HeartStart

 

FR2+ uses multiple parameters to determine if a rhythm is shockable.

 

NOTE: For safety reasons, some very low-amplitude or low-frequency rhythms

 

may not be interpreted as shockable VF rhythms. Also, some VT rhythms may not

 

be interpreted as shockable rhythms. CPR rates significantly above 100

 

compressions per minute can cause incorrect or delayed analysis by the

 

HeartStart FR2+.

 

 

Asystole

On detection of asystole, provides CPR prompt at programmed interval.

 

 

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Page 72
Image 72
Philips M3861A, M3860A manual ECG Analysis System