Setup and Advanced Mode Features

 

 

 

Setup Overview

 

 

 

The “setup” of the HeartStart FR2+ Defibrillator is made up of several

 

 

 

programmable aspects, or parameters, of FR2+ operation. Some setup

 

 

 

parameters govern specific features that are not related to the patient care

 

 

 

protocol, some are used to define the automatic patient care protocol used

 

 

 

by the FR2+, and some provide options for manual override of the protocol

 

 

 

during use.

 

 

 

 

The FR2+ comes with a factory default setup designed to meet the needs of

 

 

 

most users. If desired, your Medical Director can revise the setup. Even if no

 

 

 

changes are made, however, it is a good idea to understand the setup of your

 

 

 

FR2+ and how the different parameter settings affect the way the device

 

 

 

works.

 

 

 

 

 

Non-protocol parameters

 

 

 

The parameters listed in the following table enable features of FR2+

 

 

 

operation that are not related to the patient treatment protocol. The table

 

 

 

describes each of these non-protocol parameters, lists the settings available

 

 

 

for it, and identifies the default setting.

 

 

 

 

 

 

 

 

parameter

settings

default

description

 

 

 

 

 

 

 

 

speaker volume

 

1, 2, 3, 4,

 

8

Sets volume of the FR2+’s speaker. 1 is lowest; 8 highest.

 

 

 

5, 6, 7, 8

 

 

The speaker is used for voice prompts and the

6

 

 

 

 

 

charge-done tone.

 

 

 

 

 

 

 

 

 

 

 

 

 

record voice

 

YES, NO

 

NO

Enables or disables the audio recording during incident.

 

 

 

 

 

 

Voice recording requires use of a data card.

 

 

 

 

 

 

 

 

ECG display

 

ON, OFF

 

ON

Enables (ON) or disables (OFF) ECG display on the

 

(M3860A only)

 

 

 

 

screen.

 

 

 

 

 

 

FR2+ rhythm analysis does not require ECG display to be

 

 

 

 

 

 

on. (ECG display cannot be changed from default OFF

 

 

 

 

 

 

for M3861A.)

 

 

 

 

 

 

 

 

autosend PST

 

ON, OFF

 

OFF

Enables (ON) or disables (OFF) transmission of the

 

 

 

 

 

 

results of the FR2+’s periodic selftests (PST) from its

 

 

 

 

 

 

infrared communications port.

 

 

 

 

 

 

 

 