4-13

symptom

possible cause

 

 

Voice prompts:

• The patient is being moved or jostled.

ANALYZING INTERRUPTED

• Radio or electrical sources are

or

interfering with ECG analysis.

CANNOT ANALYZE

• The environment is dry and movement

or

around the patient is causing static

STOP ALL MOTION

electricity to interfere with ECG

 

analysis.

 

 

Voice and screen prompts:

The patient impedance is not appropriate

NO SHOCK DELIVERED

for the FR2+ to deliver a shock.

 

 

Voice prompt:

Shock has been advised but not delivered

SHOCK BUTTON NOT

within 30 seconds. (FR2+ has been

PRESSED

disarmed.)

 

 

recommended action

Stop CPR; do not touch the patient. Minimize patient motion. If the patient is being transported, stop the vehicle if needed.

Check for possible causes of radio and electrical interference and remove them from the area.

Responders and bystanders should minimize motion, particularly in dry environments that can generate static electricity.

Make sure the defibrillator pads

are correctly positioned on the

 

patient according to the diagram

4

on the back of the pads.

Make sure the defibrillator pads connector is completely inserted in the connector socket.

Press the defibrillator pads firmly to the patient's chest.

Replace the defibrillator pads if necessary.

When next prompted, press the Shock button to deliver shock.

P hili ps Me dical Sy stems

Maintaining, Testing, and Troubleshooting Your HeartStart FR2+

Page 39
Image 39
Philips M3860A, M3861A manual Symptom Possible cause