5-4

safety level

possible improper device performance

 

 

CAUTION

Use only Philips-approved data cards. The HeartStart FR2+ may perform

 

improperly if non-approved accessories are used.

 

 

 

 

safety level

possible electrical interference with ECG monitoring

 

 

WARNING

Radio-frequency (RF) interference from devices such as cellular

 

phones and two-way radios can cause improper HeartStart FR2+

 

operation. The HeartStart FR2+ should be used at least 6 feet (2

 

meters) away from RF devices, as stated in accordance with EN

 

61000-4-3:1996.

 

 

 

Defibrillation warnings and cautions

 

 

safety level

possible shock hazard

 

 

WARNING

Defibrillation current can cause operator or bystander injury. Do not

 

touch the patient during defibrillation. Disconnect the pads connector

 

from the HeartStart FR2+ before using any other defibrillator.

 

 

 

 

safety levels

possible ECG misinterpretation

 

 

WARNING

For safety reasons, some very low-amplitude or low-frequency heart

 

rhythms may not be interpreted by the HeartStart FR2+ as shockable

 

VF rhythms. Also, some VT rhythms may not be interpreted as

 

shockable rhythms.

 

 

WARNING

Handling or transporting the patient during heart rhythm analysis can

 

cause an incorrect or delayed diagnosis. If the HeartStart FR2+ gives

 

a SHOCK ADVISED prompt, keep the patient as still as possible for at

 

least 15 seconds so the HeartStart FR2+ can reconfirm the rhythm

 

analysis before a shock is delivered.

 

 

P hili ps Me dical Sy stems

M3860A/M3861A HEARTSTART FR2+

Page 46
Image 46