P hili ps Me dical Sy stems
|
|
|
| B | |
|
|
|
|
| |
|
|
|
| ||
category |
| nominal specifications |
| ||
|
|
| |||
Energy | Using adult defibrillator pads: 150 J nominal into a 50 ohm load. |
| |||
| Using infant/child |
| |||
| load. Sample pediatric energy doses: |
|
|
| |
|
| age | energy dose |
| |
|
|
|
|
|
|
|
| newborn | 14 J/kg |
| |
|
| 1 year | 5 J/kg |
| |
|
| 2 − 3 years | 4 J/kg |
| |
|
| 4 − 5 years | 3 J/kg |
| |
|
| 6 − 8 years | 2 J/kg |
| |
| Doses indicated are based on CDC growth charts for the 50th percentile weights |
| |||
| for boys.* |
|
|
|
*National Center for Health Statistics in collaboration with the National Center for Chronic Disease Prevention and Health Promotion. CDC growth charts:
Charge Control | Controlled by Patient Analysis System for automated operation. Can be |
| programmed for manual initiation using advanced mode of the M3860A. |
|
|
Charge Time | < 10 seconds typical, including confirming analysis. Charge time increases near |
from “Shock Advised” | end of battery service life. |
|
|
< 20 seconds typical, including analysis, in AED mode. | |
|
|
“Charge Complete” | Shock button flashes, audio tone sounds. |
Indicator |
|
|
|
Disarm (AED mode) | Once charged, the HeartStart FR2+ will disarm if: |
| • patient’s heart rhythm changes to |
| • a shock is not delivered within 30 seconds after the FR2+ is armed, OR |
| • the PAUSE button (if enabled) is pressed, OR |
| • the On/Off button is pressed to turn off the FR2+, OR |
• the defibrillator pads are removed from the patient or the pads connector is disconnected from the FR2+.
Technical Specifications