nominal specifications

 

 

 

 

Energy

Using adult defibrillator pads: 150 J nominal into a 50 ohm load.

 

 

Using infant/child reduced-energy defibrillator pads: 50 J nominal into a 50 ohm

 

 

load. Sample pediatric energy doses:

 

 

 

 

 

age

energy dose

 

 

 

 

 

 

 

 

 

newborn

14 J/kg

 

 

 

1 year

5 J/kg

 

 

 

2 − 3 years

4 J/kg

 

 

 

4 − 5 years

3 J/kg

 

 

 

6 − 8 years

2 J/kg

 

 

Doses indicated are based on CDC growth charts for the 50th percentile weights

 

 

for boys.*

 

 

 

*National Center for Health Statistics in collaboration with the National Center for Chronic Disease Prevention and Health Promotion. CDC growth charts: weight-for-age percentiles, revised and corrected November 28, 2000. Atlanta, GA: Centers for Disease Control and Prevention © 2000.

Charge Control

Controlled by Patient Analysis System for automated operation. Can be

 

programmed for manual initiation using advanced mode of the M3860A.

 

 

Charge Time

< 10 seconds typical, including confirming analysis. Charge time increases near

from “Shock Advised”

end of battery service life.

 

 

Shock-to-Shock Cycle Time

< 20 seconds typical, including analysis, in AED mode.

 

 

“Charge Complete”

Shock button flashes, audio tone sounds.

Indicator

 

 

 

Disarm (AED mode)

Once charged, the HeartStart FR2+ will disarm if:

 

• patient’s heart rhythm changes to non-shockable rhythm, OR

 

• a shock is not delivered within 30 seconds after the FR2+ is armed, OR

 

• the PAUSE button (if enabled) is pressed, OR

 

• the On/Off button is pressed to turn off the FR2+, OR

• the defibrillator pads are removed from the patient or the pads connector is disconnected from the FR2+.

Technical Specifications

Page 71
Image 71
Philips M3860A, M3861A manual Energy dose