P hili ps Me dical Sy stems

 

B-5

B

 

 

 

 

 

category

nominal specifications

 

 

 

 

Pacemaker Detection

On detection of a pacemaker (in advanced mode or with M3873A/M3874A FR2+

 

 

ECG assessment module), provides screen display of PACEMAKER DETECTED

 

 

alert. M3860A includes pacemaker artifact in ECG display. In both models,

 

 

pacemaker artifact is removed from the signal for rhythm analysis.

 

 

Display

 

 

 

 

 

 

category

nominal specifications

 

 

 

 

Monitored ECG Lead

ECG information is received from adult defibrillator pads in anterior-anterior (Lead

 

II)position or from FR2 infant/child reduced-energy defibrillator pads in anterior-posterior position. (Displayed on M3860A only.)

ECG information can also be displayed in the M3860A using the FR2+ ECG assessment module.

 

NOTE: The ECG display provided by the FR2+ Defibrillator is not is not

 

intended to provide diagnostic or ST segment interpretation.

 

 

Display Range

Differential: ±2 mV full scale, nominal.

(M3860A only)

 

 

 

Screen Type

High-resolution liquid crystal display (LCD) with backlight.

 

 

Screen Dimensions

2.8" wide x 2.3" high (70 mm x 58 mm).

 

 

Sweep Speed

23 mm/s nominal.

(M3860A only)

 

 

 

ECG Display

3 second-segments displayed (M3860A only).

 

 

Frequency Response

Nondiagnostic rhythm monitor 1 Hz to 20 Hz (-3 dB), nominal.

(Bandwidth)

 

 

 

Sensitivity

1.16 cm/mV, nominal.

 

 

Heart Rate Displayed

30 to 300 bpm, updated each analysis period. Displayed (M3860A only) during

(Normal Sinus Rhythm)

monitoring and advanced modes.

 

Controls and indicators

 

 

 

category

nominal specifications

 

 

LCD Screen

High-resolution, backlighted LCD screen, displays text messages and (model

 

M3860A only) ECG.

 

 

Technical Specifications

Page 73
Image 73
Philips M3860A, M3861A manual Controls and indicators