P hili ps Me dical Sy stems
symptom
Status Indicator:
SOLID RED X
Audio signal:
NONE
Status Indicator:
SOLID RED X
Audio signal:
CHIRPING
Status Indicator:
NONE
possible cause
•The battery is missing or completely depleted.
•The training & administration pack is being used in the administration function (the solid red X is normal in this case) or has been left in the FR2+ by mistake.
•A
•The training & administration pack is being used in the ADMINISTRATION function and more than 10 minutes have passed without user interaction (button press or pads change).
•The training & administration pack is being used in the TRAINING function and more than 30 minutes have passed without user interaction (button press or pads change).
The FR2+ has been physically damaged.
recommended action
•Install a new M3863A FR2 standard battery or a charged M3848A FR2+ rechargeable battery in the FR2+ and perform the battery insertion test (BIT).
•Remove the training & administration pack and install a battery.
•Remove and reinstall the battery and perform the battery insertion
rechargeable battery and repeat |
|
the test. If it fails again, do not use | 4 |
the FR2+. |
•To continue using the training & administration pack, press any button (except On/Off).
•To return the FR2+ to standby mode, remove the Pack and install a battery.
•Check for visible damage. Do not use the FR2+ if it appears to be damaged.
•Remove and reinstall the battery to perform the battery insertion
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+