4-15

P hili ps Me dical Sy stems

symptom

Status Indicator:

SOLID RED X

Audio signal:

NONE

Status Indicator:

SOLID RED X

Audio signal:

CHIRPING

Status Indicator:

NONE

possible cause

The battery is missing or completely depleted.

The training & administration pack is being used in the administration function (the solid red X is normal in this case) or has been left in the FR2+ by mistake.

A self-test detected a failure.

The training & administration pack is being used in the ADMINISTRATION function and more than 10 minutes have passed without user interaction (button press or pads change).

The training & administration pack is being used in the TRAINING function and more than 30 minutes have passed without user interaction (button press or pads change).

The FR2+ has been physically damaged.

recommended action

Install a new M3863A FR2 standard battery or a charged M3848A FR2+ rechargeable battery in the FR2+ and perform the battery insertion test (BIT).

Remove the training & administration pack and install a battery.

Remove and reinstall the battery and perform the battery insertion self-test. If it fails, install a new M3863A FR2 standard battery or a charged M3848A FR2+

rechargeable battery and repeat

 

the test. If it fails again, do not use

4

the FR2+.

To continue using the training & administration pack, press any button (except On/Off).

To return the FR2+ to standby mode, remove the Pack and install a battery.

Check for visible damage. Do not use the FR2+ if it appears to be damaged.

Remove and reinstall the battery to perform the battery insertion self-test. If it fails, install a new M3863A FR2 standard battery or a charged M3848A FR2+ rechargeable battery and repeat the test. If it fails again, do not use the FR2+.

Maintaining, Testing, and Troubleshooting Your HeartStart FR2+

Page 41
Image 41
Philips M3860A, M3861A manual FR2+ has been physically damaged