Philips M3861A Using the manual charge feature M3860A only

Models: M3860A M3861A

1 102
Download 102 pages 22.88 Kb
Page 59
Image 59

P hili ps Me dical Sy stems

6-11

In the M3861A HeartStart FR2+, the patient’s ECG is not displayed; in the M3860A, the display includes the patient’s ECG and heart rate.

Press the lower Option button (ANALYZE) to initiate rhythm analysis by the FR2+. If a shock is advised, the FR2+ automatically charges, and prompts you to press the Shock button.

After shock delivery, the HeartStart FR2+ returns to the advanced mode display and monitors the patient’s heart rhythm. If a potentially shockable rhythm is detected, the text and voice prompts advise you to PRESS ANALYZE.

N O T E : If you do not press the lower Option button (labeled ANALYZE) to initiate rhythm analysis when prompted, the HeartStart FR2+ does not analyze and prompt if a shock is advised. It is important that you understand that entering the advanced mode entails taking responsibility for these functions.

If the rhythm analysis results in a Shock Advised decision, the FR2+ begins charging, prompts you to press the Shock button, and displays a MANUAL DISARM option at the top of the screen. If for any reason you want to cancel the shock, press the upper Option button to disarm the FR2+.

To return to non-manual, AED mode operation, turn the FR2+ off by pressing the On/Off button. Then turn the FR2+ on by pressing the On/Off button again.

Using the manual charge feature (M3860A only)

The manual charge feature is available only in the M3860A, when enabled in

6

 

setup.

 

To enter the advanced mode during use of an FR2+ that has this feature

 

enabled, press both Option buttons simultaneously. This brings up a screen

 

that includes a highlighted line at the top, labeled MANUAL, with an

 

arrowhead pointing to the upper Option button, and another at the bottom,

 

labeled ANALYZE, with an arrowhead pointing to the lower Option button.

 

When the advanced mode is entered, display of the patient’s ECG and heart

 

rate is automatically initiated.

 

Pressing the lower Option button (ANALYZE) provides user-initiated rhythm

 

analysis as described above. Pressing the upper Option button (MANUAL)

 

brings up a new screen.

 

Setup and Advanced Mode Features

Page 59
Image 59
Philips M3861A Using the manual charge feature M3860A only